Multi-Site Real-World Performance of Commercial AI for Pulmonary and Incidental Pulmonary Embolism Detection
Tuesday, September 29, 2026
Pulmonary embolism (PE) is a leading cause of cardiovascular mortality, yet the real-world performance of FDA-cleared AI detection models remains incompletely characterized. We retrospectively evaluated two FDA-cleared AI algorithms from a single commercial platform (Aidoc Medical BriefCase), one for PE triage on dedicated CT pulmonary angiography (CTPA; n = 30,678) and one for incidental PE (iPE) detection on routine contrast-enhanced CTs (n = 37,191), across a 17-facility academic health system. Reference-standard labels were extracted from radiology reports using a validated LLM pipeline (97% accuracy, kappa = 0.94). The PE model achieved 86.8% sensitivity and 99.1% specificity, with sensitivity declining from 99.3% for saddle emboli to 72.9% for subsegmental PE, and from 89.7% for acute to 65.3% for non-acute PE. The iPE model achieved 73.5% sensitivity and 99.8% specificity. Both models demonstrated lower sensitivity than FDA-clearance benchmarks while exceeding cleared specificity, with diminishing performance for peripheral and non-acute emboli mirroring known human reader limitations and underscoring the need for standardized post-market surveillance of AI-enabled medical devices.
Daily healthcare AI brief
Get the free healthcare AI briefing
One concise briefing built from reporting, research, policy, blogs, videos, and full podcast transcripts.
Free. No account needed. Unsubscribe anytime. Already subscribed? Sign in to save and personalize.